Polyvinyl Alcohol (PVA) polymer packed aqueous SEC (size-exclusion chromatography) column can be used with a wide range of solvents including water to nonpolar solvents. Depends on the solvent polarity, JAIGEL-GS can be GFC (gel-filtration chromatography) or adsorption and partition modes.
Standard packing solvent: water
Flow rate for the analytical size column is typically 0.4-0.6mL/min, max 1.0mL/min
Pre-column size is 7.6x50mm and denoted with "-P" after the column model numbers.
JAIGEL-GS Column Size, Exclusion Limit, and Resolution
Items
Exclusion Limit (Pullulan)
Preparative Columns
Analytical
pH Range
Pre-Column
Applications
I.D. 20mm
I.D. 7.6mm
Length 500mm
Length 300mm
Length 300mm
JAIGEL-GS220
3,000
GS220 (R>12,000)
-F (R>7,000)
-A (R>16,000)
2-9
-P2
peptides, water-solble oligomers, small saccharides
JAIGEL-GS310
40,000
GS310 (R>12,000)
-F (R>7,000)
-A (R>16,000)
2-9
-P1
peptides, steroids
JAIGEL-GS320
40,000
GS320 (R>12,000)
-F (R>7,000)
-A (R>16,000)
2-12
-P2
nucleic acids,peptides, saccharies, hormones
JAIGEL-GS510
3x105
GS510 (R>12,000)
-F (R>7,000)
-A (R>16,000)
2-9
-P1
hydrophobic polymers, proteins, nucleotides
JAIGEL-GS520
3x105
GS520 (R>12,000)
-F (R>7,000)
-A (R>15,000)
2-12
-P2
proteins in blood, oligosaccharides, albumin, nucleotides